Exam 39: Automation in Hematology

arrow
  • Select Tags
search iconSearch Question
flashcardsStudy Flashcards
  • Select Tags

Which of the following analyzers uses radio frequency to detect and differentiate leukocytes?

Free
(Multiple Choice)
4.8/5
(81)
Correct Answer:
Verified

C

Explain the purpose of the following used in the standard hematologic cell-counting detection methods: a.VCS flow cell b.Photodetector c.Sheath fluid d.Direct current detection block

Free
(Essay)
4.9/5
(66)
Correct Answer:
Verified

VCS flow cell: to detect each WBC’s volume, conductivity, and scatter for determination of the five part differential
Photodetector: to measure the absorbance of the unknown blood sample passing through it (used in spectrophotometric detection of Hgb)
Sheath fluid: to promote hydrodynamic focusing of cells passing through photodetector
Direct current detection block: to detect electrical resistance when cells pass through the aperture

Which of the following is measured immunologically on the Diagnostica Stago STA Compact?

Free
(Multiple Choice)
4.8/5
(75)
Correct Answer:
Verified

A

An analyzer is measuring a clot forming by the interaction of a stationary probe with a moving probe in a continuous motion.Suddenly the moving probe stops,and a clot is seen.Name the analyzer.

(Multiple Choice)
5.0/5
(86)

The immature platelet fraction (IPF)is a parameter derived from the Sysmex XE-2100's RET scattergram.An increased IPF is associated with:

(Multiple Choice)
4.9/5
(71)

Hydrodynamic focusing is used in automated cell-counting instruments to:

(Multiple Choice)
4.8/5
(87)

Name two analytes that can be measured immunologically,and list two instruments that currently use this method.

(Essay)
4.9/5
(74)

Using the Fibrometer instrument,clot detection occurs when:

(Multiple Choice)
4.9/5
(75)

Which of the following analyzers measures reticulocytes by sphering the blood cells prior to analysis?

(Multiple Choice)
4.8/5
(78)

Using the chromogenic methodology,a coagulation protein's activity level is related to:

(Multiple Choice)
4.9/5
(85)

Which of the following represents the degree of light scatter that reflects cell complexity on the CELL-DYN Sapphire?

(Multiple Choice)
4.9/5
(79)

Which of the following instruments does not rely on impedance to enumerate cells:

(Multiple Choice)
4.8/5
(70)

The purpose of the photodetector on the Diagnostica Stago STA Compact is to:

(Multiple Choice)
4.8/5
(93)

Which analyzer has the ability to quantify nucleated RBCs and calculate a corrected WBC count?

(Multiple Choice)
4.9/5
(80)

Which technologies are used to generate the LH 750's WBC scatterplot?

(Multiple Choice)
5.0/5
(92)

Cell volume and cell number are required elements for which of the following?

(Multiple Choice)
4.8/5
(86)

Which of the following binds to residual RNA in reticulocytes on the Sysmex XE-2100 analyzer?

(Multiple Choice)
4.8/5
(82)

The immunologic technique available on the AMAX Destiny Plus utilizes specific reagents with:

(Multiple Choice)
4.8/5
(78)

Choose the correct statement regarding chromogenic analysis in hemostasis testing from the following choices.

(Multiple Choice)
5.0/5
(77)

Which of the following parameters are calculated on the Coulter Gen-S?

(Multiple Choice)
4.9/5
(90)
Showing 1 - 20 of 38
close modal

Filters

  • Essay(0)
  • Multiple Choice(0)
  • Short Answer(0)
  • True False(0)
  • Matching(0)